2.2. Modes 765 Dosimat 6 2.1.1 Key <RATE> The inquiries of this key are identical for all modes. Expelling and filling rate This key is

Strona 3 - Content

2.2 Modes 765 Dosimat 7 Example: Selection of mode "DIS C", cumulative dispensing. Press <MODE>. Display shows that mode whic

Strona 4

2.2. Modes 765 Dosimat 8 Calculation values b = 0. ml Blank value Input range: 0...±999.999 mL f = 1. Factor Input ra

Strona 5

2.2 Modes 765 Dosimat 9 2.2.2 Mode DIS R, Repetitive Dispensing Dosimat is dosing a stored dispensing volume if <GO> is pressed. The burett

Strona 6 - 1 Overview

2.2. Modes 765 Dosimat 10 2.2.4 Mode PIP, Pipetting Aspirating and subsequent expelling of a stored pipetting volume. V-PIP 0.1 ml Pipe

Strona 7 - Rear view of instrument:

2.2 Modes 765 Dosimat 11 • A new air bubble is built with every preparation step ("prep.") e.g. its volume increases. If you wish to kee

Strona 8

2.2. Modes 765 Dosimat 12 Sequence of DIL Standard mode DIL. Preparation step: Hold burette tip free at working height. Ready to aspirate th

Strona 9

2.2 Modes 765 Dosimat 13 Content entries which can be implemented in the CNT D mode are summarized below and designated with a bold frame Concentr

Strona 10 - 2.2 Modes

2.2. Modes 765 Dosimat 14 The formulae for calculation of the volume to be dispensed "add V" are shown in the following table, with

Strona 11 - 2.2.1 Mode DOS, Dosing

2.2 Modes 765 Dosimat 15 The following table shows factors for the most common ionic standards: Cation Standard prepared from: Factor f Anion

Strona 13

2.2. Modes 765 Dosimat 16 Since the volume of the solvent V0 is dispensed in the operational method of the CNT D mode, higher concentrations req

Strona 14 - 2.2.4 Mode PIP, Pipetting

2.3 User memory 765 Dosimat 17 2.3 User memory Up to 10 modes, complete with their user selected specific parameters, can be stored in the user me

Strona 15 - 2.2.5 Mode DIL, Diluting

2.4. Special settings 765 Dosimat 18 2.4 Special settings Special settings can be executed by pressing keys <0> and <—> simultaneou

Strona 16 - DIL ↓ prep

3.1 Special messages and error messages 765 Dosimat 19 3 Error messages, troubleshooting blinking value The value keyed in is out of the input ran

Strona 17 - Specified

3.1. Special messages and error messages 765 Dosimat 20 volume <resol.! The volume to be expelled is smaller than the resolution of the bu

Strona 18

3.2 Diagnosis 765 Dosimat 21 3.2 Diagnosis 3.2.1 General The 765 Dosimat is a very precise and dependable feeding instrument. Thanks to its rugged

Strona 19

3.2. Diagnosis 765 Dosimat 22 3.2.2 Summary 3.2.4 Diagnosis of cylinder code...23 3.2.5 Diagnosis of

Strona 20

3.2 Diagnosis 765 Dosimat 23 3.2.4 Diagnosis of cylinder code 1.Prepare instrument for diagnostic test (see chap. 3.2.3). 2.<0> cylinder co

Strona 21 - 2.3 User memory

3.2. Diagnosis 765 Dosimat 24 2.<2> display test 3.<GO> Characters are generated for an optical check of the display: 1. The d

Strona 22 - 2.4 Special settings

3.2 Diagnosis 765 Dosimat 25 1.Prepare instrument for diagnostic test (see chap. 3.2.3). 2.Connect voltmeter, DVM or recorder by means of cable 3.9

Strona 23

Content 765 Dosimat Content 1 Overview ...

Strona 24

3.2. Diagnosis 765 Dosimat 26 1.Prepare instrument for diagnostic test (see chap. 3.2.3). 2.<4> timer dig. test 3.<GO> timer dig.

Strona 25 - Equipment required:

3.2 Diagnosis 765 Dosimat 27 5.<CLEAR> diagn. key 0...9 6.Remove test plug 3.496.8360. 3.2.11 RAM-test 1.Prepare instrument for diagnostic t

Strona 26 - 3.2.2 Summary

3.2. Diagnosis 765 Dosimat 28 3.Power off and wait for 5 s. 4.Power on and simultaneously press key <0> and keep pressed until: special k

Strona 27 - 3.2.6 Diagnosis of display

3.2 Diagnosis 765 Dosimat 29 The spindle moves 80 mm with respect to spindle zero. Instead of the spindle height one can also measure the expelled

Strona 28

3.3. RAM-initialisation 765 Dosimat 30 3.3 RAM-initialisation In rare cases, it is possible that major interference signals such as line spikes

Strona 29

3.4 Releasing a locked spindle 765 Dosimat 31 3.4 Releasing a locked spindle with inserted Exchange Unit The burette drive may very occasionally j

Strona 31 - 3.2.11 RAM-test

4.1 General 765 Dosimat 33 4 Operation via RS232 Interface 4.1 General The Dosimat offers an extensive remote control. Data transmission occurs

Strona 32

4.2. Control commands 765 Dosimat 34 4.2 Control commands Command Explanation Live Notes REMOTE ON Remote control on Y 4.2.1 REMOTE OFF

Strona 33

4.2 Control commands 765 Dosimat 35 Command Explanation Live Notes AFILL ON Auto fill on Y 4.2.33 AFILL OFF Auto fill off Y 4.2.34 QDISPL

Strona 35 - On no account touch these!

4.2. Control commands 765 Dosimat 36 4.2.3 G GO, not live, 1-byte command 'GO' triggers dosing in all modes. In mode DOS dosing goe

Strona 36

4.2 Control commands 765 Dosimat 37 Information byte 2: Bit Function 0 1= Wrong command code 1 1 = Parameter corrected to its limit value 2 1 =

Strona 37 - 4.1 General

4.2. Control commands 765 Dosimat 38 4.2.15 MDC Mode DIS C with previous parameters, not live Selection of mode DIS C without changing the act

Strona 38 - 4.2 Control commands

4.2 Control commands 765 Dosimat 39 4.2.22 PSM VALUE Parameter 'smpl', live Value = -1E33 ... -1E-37 , 0 , 1E-37 ... 1E33 Setting sampl

Strona 39

4.2. Control commands 765 Dosimat 40 Cylinder ml rate min ml/min rate max ml/min 1 0.001 3 5 0.005 15 10 0.010 30 20 0.020 60 50 0.050 150 If

Strona 40

4.2 Control commands 765 Dosimat 41 The entry is automatically corrected to a multiple of 1/10 000 V(B) depending on the volume of the cylinder: Cy

Strona 41

4.2. Control commands 765 Dosimat 42 Cylinder ml Volume min ml/min Volume max ml/min 1 0.001 999.999 5 0.001 999.999 10 0.001 999.999 20 0.0

Strona 42

4.2 Control commands 765 Dosimat 43 The binary value is transmitted in 4 bytes, where only the low-order nibble of a byte contains a 4 bit informat

Strona 43

4.2. Control commands 765 Dosimat 44 4.2.46 QAD Query rate down 'analogue on/off', live Request to send the information rate down

Strona 44

4.3 Handshake and other properties 765 Dosimat 45 4.3 Handshake and other properties 4.3.1 Handshake full The arrows show the direction of the si

Strona 45

Overview 765 Dosimat 1 Explanation of symbols: < > means "key", e.g. <GO> means key "GO" DOS...0.00

Strona 46

4.3. Handshake and other properties 765 Dosimat 46 4.3.3 General properties of the RS 232 interface The Dosimat is configured as DTE (Data Term

Strona 47

4.4 Pin assignment of the RS232 / I/O socket 765 Dosimat 47 tp ca. 200 us tp ca.150 ms tp tp tp ≥ 1 ms tp tp ≥ 1 ms tp 4.4 Pin assignment of the R

Strona 48

4.4. Pin assignment of the RS232 / I/O socket 765 Dosimat 48 4.4.2 RS232 interface, 25 and 8 pin sockets Transmitted Data (TxD). If no

Strona 49 - 4.3.2 Handshake none

4.4 Pin assignment of the RS232 / I/O socket 765 Dosimat 49 Definition of signal states - Data interchange circuits (TxD, RxD) voltage negative

Strona 50

4.4. Pin assignment of the RS232 / I/O socket 765 Dosimat 50

Strona 51 - 4.4.1 I/O socket, 25 pins

5.1 Technical specifications 765 Dosimat 51 5 Appendix 5.1 Technical specifications Exchange units 1, 5, 10, 20, 50 mL burette cylinder volumes,

Strona 52 - 765 Dosimat

5.1. Technical specifications 765 Dosimat 52 Display LCD, 16 characters Size of characters: 4.84 x 8.01 mm Material Cabinet Polybutylene t

Strona 53

5.2 Warranty and certificates 765 Dosimat 53 5.2 Warranty and certificates 5.2.1 Warranty The warranty regarding our products is limited to rectif

Strona 54

5.2. Warranty and certificates 765 Dosimat 54 5.2.2 Certificate of Conformity and System Validation This is to certify the conformity to the s

Strona 55 - 5 Appendix

5.2 Warranty and certificates 765 Dosimat 55 Ionenanalytik • Analyse des ions • Ion analysis • Análisis iónico 765 Dosimat EU- Declaration o

Strona 56

Overview 765 Dosimat 2 1 Overview Front view of instrument: 1 Exchange unit Normally the models with automatic cock changeover. Note: Choose

Strona 57 - 5.2.1 Warranty

5.3. Connection of a balance 765 Dosimat 56 5.3 Connection of a balance The balance is connected to the RS232 output (6) of the Dosimat. Conn

Strona 58

5.5 Continuous dosing with two Dosimats 765 Dosimat 57 5.5 Continuous dosing with two Dosimats Two 765 Dosimats are suitable for continuous dosin

Strona 59

5.5. Continuous dosing with two Dosimats 765 Dosimat 58 Example: Continuous dosing of 55 mL. Exchange units with volumes of 20 mL and 10 mL.

Strona 60 - 5.4 Connection of a printer

5.6 Scope of delivery and ordering designations 765 Dosimat 59 5.6 Scope of delivery and ordering designations Dosimat 765 ...

Strona 61

5.6. Scope of delivery and ordering designations 765 Dosimat 60 Options Accessories to separate order and on payment of extra charge: 806 Excha

Strona 62

Index 765 Dosimat 61 Index Keys are marked with < >, display texts are in bold characters, and pages concerning the green part are printed in

Strona 63

Index 765 Dosimat 62 M M ... 13 <MODE> ... 6 Mode Lo

Strona 64

Overview 765 Dosimat 3 Rear view of instrument: 6 Data inputs and outputs Via data transfer interface according to RS 232 C including optiona

Strona 65

2.1. Keypad 765 Dosimat 4 2 Operation with the keyboard 2.1 Keypad 6.2149.000 SMPL 9 BLANK 7 FACTO

Strona 66

2.1 Keypad 765 Dosimat 5 Rules for data input: • On entering a negative number, key in minus sign first; <-> is not a "change of sign