4.2. Diagnosis
794 Basic Titrino
145
RS232 test-connector?
Insert the 3.496.8480 test plug in 'RS 232' port.
Press <Enter>.
Test sequence:
1. In case of an error an error message is displayed. If for example no test plug is con-
nected, error message
error 68
appears).
2. If no error is found, after about 5 s display shows
RS232 test o.k.
Remove test plug.
Press <Clear>.
diagnose press key 0...9
10 Spindle drive and stopcock changeover
Press <Clear>.
The Titrino fills (only if an exchange unit is inserted)
The Titrino gets out of the diagnosis menu and returns to the user program.
Remove exchange unit.
Check spindle zero position, see Fig. 5.
0.2 - 0.6 mm
Fig. 5 Fig. 6
The spindle must be 0.2 - 0.6 mm below the edge of the sliding plate.
The bar of the stopcock coupling must be exactly parallel to the lateral edges of the MPT
Titrino, see Fig.6.
Reinsert Exchange Unit.
Titrino fills.
The display of before reappears.
Knob 'dV/dt' to right stop.
Press the <DOS> key (on instrument) until the piston rod reaches the top and at the
same time measure the time from start to end.
XXX X ********
cylinder em
p
ty !
mind selected language!
Przeglądanie stron 148