Metrohm Vision – Tutorial Instrukcja Użytkownika Strona 26

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 40
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 25
1.6 Multilinear Regression 
NIRS Vision Tutorial Operation  19
Perform Sample Selection
1. Click on Select Apply Math.
2. In the Edit Sample Selection Parameters dialog box enter the
following settings:
Selection Method Mahalanobis Distance in Principal Component
Space
Math Treatment 2nd derivative
Leave all other settings unchanged
3. Click on [OK].
4. In the Edit Math Properties dialog box in the Principal Components
ID Parameters tab set the following wavelength regions:
Region 1
Minimum: 416
Maximum: 1080
Region 2
Minimum: 1120
Maximum: 2484
Cum Var. 95 %
Click on [OK]
5. In the Edit Math Properties dialog box in the 2nd derivative tab set
the following values:
Segment: 10
Gap: 0
6. In the Sample Selection Options dialog box apply or confirm the
following settings:
Find Outliers enabled
Outlier Threshold 0.95
Threshold Type Probability Level
Find Redundant enabled
By number of samples enabled
Training Set 75 %
Acceptance Set 25 %
Click on [OK]
Vision now performs calculations according to the parameters defined in the
previous steps.
Przeglądanie stron 25
1 2 ... 21 22 23 24 25 26 27 28 29 30 31 ... 39 40

Komentarze do niniejszej Instrukcji

Brak uwag