Metrohm Inline Dialysis Instrukcja Użytkownika Strona 24

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 30
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 23
4.1 Operation
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
IC equipment: Dialysis
2
Measuring the standard solution with dialysis
Set the dialysis time in the software to 5 minutes.
Immerse the sample aspiration capillary in the standard solution
used in step 1.
Start a determination in the software and wait until the chroma-
togram has been evaluated.
3
Additional measurements
Increase the dialysis time in the software by 1 minute each time
and start the determination.
Repeat the measurements until the measured values are stable.
4
Determining the optimum dialysis time
The recovery rates can be calculated from the result without dialy-
sis and the results with dialysis. You can determine the optimum
dialysis time by recording the recovery rate in relation to the dialy-
sis time.
4.1.2 Recommended procedure for the dialysis
For sample determination with dialysis, it is best to proceed according to
the following sequence:
Carrying out a dialysis
1
Preparing the acceptor solution
To avoid disruptions caused by air bubbles in the acceptor flow,
degas the ultrapure water used as acceptor solution for at least
10 minutes.
2
Preparing the sample
To avoid blockage of the sample channel, always centrifuge sam-
ples with a high content of suspended or solid particles in a tab-
letop centrifuge for 5 minutes at 10,000 rpm.
3
Putting the IC system into operation
Switch on all required devices.
Start the software and load the required IC system.
Condition the IC system.
Przeglądanie stron 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 30

Komentarze do niniejszej Instrukcji

Brak uwag