5.2 Maintenance
■■■■■■■■■■■■■■■■■■■■■■
18
■■■■■■■■
IC Equipment Thin-Layer cell
■ The working electrode has been removed from the measuring cell.
1
Note
When cleaning the auxiliary electrode, the edges of the spacer
may get damaged.
The spacer must then be replaced.
■ Moisten a cotton swab with 2 mol/L of nitric acid.
■ Use this to carefully wipe the auxiliary electrode.
The edges of the spacer may be damaged when wiping the auxili-
ary electrode.
If the auxiliary electrode is tenaciously stained, we recommend
that the spacer be removed prior to cleaning the auxiliary elec-
trode (see Chapter 5.2.3, page 15).
2
Rinse off the measuring cell (except for the working electrode and
the reference electrode) under running water and dry with a lint-free
cloth.
Note
When you rinse off the measuring part, take care to ensure that
the chip in the cell holder does not get wet.
Cleaning the working electrode
1
Clean the working electrode in accordance with the directions con-
tained in the leaflet for the working electrode or polish it if necessary.
2
Reinsert the working electrode (see Chapter 3.1, page 4).