Metrohm 797 VA Computrace Instrukcja Użytkownika Strona 250

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 249
8 How to ...?
797 VA Computrace – Software
238
The determination tab of the window for CVS:
Note: To make sure that always the latest calibration file is ta-
ken for the calculation, the name of the calibration file defined
for the parameter Calibration curve on the Determination tab (of
the window EDIT WORKING METHOD PARAMETERS) with cali-
bration
"DT Suppressors with calibration curve", must match
with the name defined for the parameter Sample identifier on
the
Determination tab (of the window EDIT WORKING METHOD
PARAMETERS
) with Calibration technique "DT Record calibration
curve".