7.2 Passport
IC Net 2.0
101
Run Number of the current run starting from the
very first one. All runs are automatically
numbered by the system (read-only).
To reset the Run number press [Ctrl] [F8].
The Analysis number window is opened
where the run number for the next run can
be modified to the desired value.
Batch X/Y: total number of started sample queues
/ current analysis number in the sample
queue (read-only).
7.2.2 Sample
Tab Sample of the passport with sample information.
Info 1 First sample description (max. 256 char-
acters).
Info 2 Second sample description (max. 256
characters).
Volume Injected volume in µL.
Dilution Dilution of the sample.
Vial number Autosampler vial position to take sample
from.
Amount Sample amount. If this value is different for
the calibration run (c) and the sample run
(s), the component concentrations of the
sample are calculated as follows:
C
s
= C
c
Amount
s
/ Amount
c