Metrohm IC Net 2.0 Instrukcja Użytkownika Strona 36

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 196
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 35
4 Systems
IC Net 2.0
28
E295
Storage of memory values failed (repeat
last action; if error reappears, call Metrohm
service).
E296
Instrument stopped (restart instrument; if
error reappears, call Metrohm service).
E297
Storage of remote line values failed (repeat
last action; if error reappears, call Metrohm
service).
E298
Storage of flow correction value failed
(repeat last action; if error reappears, call
Metrohm service).
E299
Storage of break time values failed (repeat
last action; if error reappears, call Metrohm
service).
E300
High-pressure pump faulty (restart pump; if
error reappears, call Metrohm service).
E301
Injection valve blocked (check injection
valve; if error reappears, call Metrohm
service).
E302
Suppressor module blocked (check sup-
pressor; if error reappears, call Metrohm
service).
E303
Storage of maintenance information failed
(repeat last action; if error reappears, call
Metrohm service).