6 Devices
IC Net 2.3
230
Click button <Add> to open the ADD NEW RACK DATA window
and add a new rack.
Settings
Rotate angle Definition of the rotate angle needed for the
command Move +/- rotate.
Entry range: 0..270 °, default 5 °
Swing angle
Definition of the swing angle needed for the
command Move +/- swing.
Entry range: 0..180 °, default 5 °
Mixing volume
Definition of the total solution volume after
dilution.
Entry range: 0.001…1000 mL, default 10 mL
Reagent volume
Definition of the volume of the reagent
added.
Entry range: 0.001…Mixing volume mL, de-
fault
0 mL
Peristaltic pump speed
Definition of the speed of the peristaltic
pump.
Entry range:
-15, …, -1, 1, …, 15, default 15
<Set>
Send parameters to 838 Advanced Sample
Processor.