Metrohm IC Net 2.3 Instrukcja Użytkownika Strona 316

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 389
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 315
7 Methods
IC Net 2.3
306
Name Channel’s name. This text appears as
Channel label in the chromatogram.
Units User-defined units of detector response.
Input Number of data acquisition channel.
Minimum Minimum value of the linear range of the AD
converter (in ADC conversion units). This
parameter is used to detect an underflow
condition.
Zero A signal level on a baseline (in ADC con-
version units). This parameter is a result of
the ADC calibration.
Maximum Maximum value of the linear range of the
AD interface (in ADC conversion units). This
parameter is used to detect overflow condi-
tion.
Range An input signal value in maximum in user-
defined units.
Range = (Maximum - Zero) Coef
Coefficient ADC sensitivity coefficient (weight of ADC
bit in user-defined units).
Noise Estimated baseline noise value of the
channel in ADC conversion units (bits).
Shift Shift of the current channel in time. This pa-
rameter is useful in the case of multi-
channel chromatograms only.
(multi)Channel ratio Spectral ratios. The option is valid for multi-
channel chromatograms only.
Przeglądanie stron 315
1 2 ... 311 312 313 314 315 316 317 318 319 320 321 ... 388 389

Komentarze do niniejszej Instrukcji

Brak uwag